| Regarding U.S. Surgical (U.S.S.) staple and suture compatibility with Magnetic Resonance Imaging (MRI) and Nuclear Magnetic Resonance (NMR) procedures with static fields of 3.0 Tesla or less; Titanium, as found in the majority of U.S.S. metallic staples or the ChemoSite* ports from U.S.S., is acceptable for use with MRI and NMR procedures as it is nonferromagnetic. U.S.S. 316L stainless steel staples, as in the Royal*, Appose*, Concorde*, Signet*, and MultiFire Premium* skin staplers, are compatible with MRI and NMR procedures with static magnetic fields of 3.0 Tesla or less. The stainless steel staples have a very low mass and their composition is considered non-magnetic. They will not vibrate or heat up, although there may be slight image distortion. Other U.S.S. non-magnetic stainless steel staples compatible with MRI/NMR procedures can be found in the GIA* 50 & 90 Premium and PI* Single Use Loading Units. This also applies to U.S.S. Stainless Steel Sutures and Flexon* Cardiac Pacing Leads. Literature supports that passive implants such as the U.S.S. staples and sutures are acceptable to undergo magnetic resonance procedures immediately after implantation at forces of 3.0 Tesla or less. Please note that this does not apply to the following;
Do not perform magnetic resonance procedures if tissue integrity and implant's ability to remain properly attached are of concern, or if proper identification of the implant location cannot be obtained. For more information, please contact: Michael Ball, Manager
|